ARUP
TSP - Total Suspended Particulates High Volume Sampler
In-situ Calibration Report
 
Calibration Date 4-Feb-06 Next Calibration Date 4-Apr-06
Station ASR1 Equipment no. E.HVS.01
    Ambient Condition    
Temperature, Ta (K) 288.7   Pressure, Pa (mmHg) 770.2
    Orifice Transfer Standard Information    
Equipment no. E.CAL.01      
Slope, mo 1.55569 Intercept, co -0.00495
Last Calibration Date 6-Jul-05   Next Calibration Date 6-Jul-06
  mo x Qstd + co = [∆O x (Pa/760) x (298/Ta)]1/2     
     Qstd = {[∆O x (Pa/760) x (298/Ta)]1/2 - co} / mo    
Calibration Point Orifice Manometer   Orifice Qstd (CMM) HVS Manometer  [∆H x (Pa/760) x (298/Ta)]1/2
Reading, ∆O (inch) x-axis Reading, ∆H (inch) y-axis
1 7.1 1.75 7.3 2.76
2 6.0 1.61 6.1 2.53
3 5.1 1.49 5.0 2.29
4 4.1 1.33 4.2 2.10
5 3.0 1.14 3.1 1.80
By Liner Regression of y on x
Slope, mh = 1.5589 Intercept, ch = 0.0084
*Correction Coefficient, R =  0.9980
Calibration Result: ACCEPT
* If the Correlation Coefficient, R  is < 0.9900. Checking and Recalibration are require.
Remark:  ______________________________________________________________________________________________
Calibrated By:  ______________________________ Date:  _______________________
Checked By:  _______________________________ Date:  _______________________
8/7/2000
ARUP
TSP - Total Suspended Particulates High Volume Sampler
In-situ Calibration Report
 
Calibration Date 4-Apr-06 Next Calibration Date 4-Jun-06
Station ASR1 Equipment no. E.HVS.01
    Ambient Condition    
Temperature, Ta (K) 299.1   Pressure, Pa (mmHg) 758.2
    Orifice Transfer Standard Information    
Equipment no. E.CAL.01      
Slope, mo 1.55569 Intercept, co -0.00495
Last Calibration Date 6-Jul-05   Next Calibration Date 6-Jul-06
  mo x Qstd + co = [∆O x (Pa/760) x (298/Ta)]1/2     
     Qstd = {[∆O x (Pa/760) x (298/Ta)]1/2 - co} / mo    
Calibration Point Orifice Manometer   Orifice Qstd (CMM) HVS Manometer  [∆H x (Pa/760) x (298/Ta)]1/2
Reading, ∆O (inch) x-axis Reading, ∆H (inch) y-axis
1 7.0 1.70 7.2 2.68
2 6.1 1.59 6.3 2.50
3 5.0 1.44 5.2 2.27
4 4.0 1.28 4.2 2.04
5 3.1 1.13 3.1 1.76
By Liner Regression of y on x
Slope, mh = 1.6034 Intercept, ch = -0.0389
*Correction Coefficient, R =  0.9990
Calibration Result: ACCEPT
* If the Correlation Coefficient, R  is < 0.9900. Checking and Recalibration are require.
Remark:  _______________________________________________________________________________________________________________________
Calibrated By:  ______________________________ Date:  ____________________
Checked By:  _______________________________ Date:  _____________________
8/7/2000
ARUP TSP - Total Suspended Particulates High Volume Sampler
In-situ Calibration Report
 
Calibration Date 16-Mar-06 Next Calibration Date 16-May-06
Station ASR2 Equipment no. E.HVS.02
    Ambient Condition    
Temperature, Ta (K) 293.7   Pressure, Pa (mmHg) 760.0
    Orifice Transfer Standard Information    
Equipment no. E.CAL.01  
Slope, mo 1.55569 Intercept, co -0.00495
Last Calibration Date 6-Jul-05   Next Calibration Date 6-Jul-06
  mo x Qstd + co = [∆O x (Pa/760) x (298/Ta)]1/2     
     Qstd = {[∆O x (Pa/760) x (298/Ta)]1/2 - co} / mo    
Calibration Point Orifice Manometer   Orifice Qstd (CMM) HVS Manometer  [∆H x (Pa/760) x (298/Ta)]1/2
Reading, ∆O (inch) x-axis Reading, ∆H (inch) y-axis
1 7.0 1.72 7.4 2.74
2 5.9 1.58 6.5 2.57
3 4.9 1.44 5.3 2.32
4 4.0 1.30 4.6 2.16
5 2.9 1.11 3.2 1.80
By Liner Regression of y on x
Slope, mh = 1.5305 Intercept, ch = 0.1346
*Correction Coefficient, R =  0.9969
Calibration Result: ACCEPT
* If the Correlation Coefficient, R  is < 0.9900. Checking and Recalibration are require.
Remark:  ______________________________________________________________________________________________
Calibrated By:  ______________________________ Date:  ______________________________
Checked By:  _______________________________ Date:  ______________________________
8/7/2000
ARUP TSP - Total Suspended Particulates High Volume Sampler
In-situ Calibration Report
 
Calibration Date 1-Mar-06 Next Calibration Date 1-May-06
Station ASR3 Equipment no. E.HVS.01
    Ambient Condition    
Temperature, Ta (K) 282.3   Pressure, Pa (mmHg) 766.0
    Orifice Transfer Standard Information    
Equipment no. E.CAL.01      
Slope, mo 1.55569 Intercept, co -0.00495
Last Calibration Date 6-Jul-05   Next Calibration Date 6-Jul-06
  mo x Qstd + co = [∆O x (Pa/760) x (298/Ta)]1/2     
     Qstd = {[∆O x (Pa/760) x (298/Ta)]1/2 - co} / mo    
Calibration Point Orifice Manometer   Orifice Qstd (CMM) HVS Manometer  [∆H x (Pa/760) x (298/Ta)]1/2
Reading, ∆O (inch) x-axis Reading, ∆H (inch) y-axis
1 5.9 1.61 6.3 2.59
2 5.0 1.49 5.3 2.37
3 4.0 1.33 4.5 2.19
4 3.0 1.15 3.5 1.93
5 2.5 1.05 3.2 1.85
By Liner Regression of y on x
Slope, mh = 1.3283 Intercept, ch = 0.4235
*Correction Coefficient, R =  0.9972
Calibration Result: ACCEPT
* If the Correlation Coefficient, R  is < 0.9900. Checking and Recalibration are require.
Remark:  ______________________________________________________________________________________________
Calibrated By:  ______________________________ Date:  ______________________________
Checked By:  _______________________________ Date:  ______________________________
8/7/2000