TSP - Total Suspended Particulates High Volume Sampler
In-situ Calibration Report (ASR1)
 
Calibration Date 3-Jun-06 Next Calibration Date 3-Aug-06
Station ASR1 Equipment no. E.HVS.01
    Ambient Condition    
Temperature, Ta (K) 300.7   Pressure, Pa (mmHg) 756.1
    Orifice Transfer Standard Information    
Equipment no. E.CAL.01      
Slope, mo 1.55569 Intercept, co -0.00495
Last Calibration Date 6-Jul-05   Next Calibration Date 6-Jul-06
  mo x Qstd + co = [∆O x (Pa/760) x (298/Ta)]1/2     
     Qstd = {[∆O x (Pa/760) x (298/Ta)]1/2 - co} / mo    
Calibration Point Orifice Manometer   Orifice Qstd (CMM) HVS Manometer  [∆H x (Pa/760) x (298/Ta)]1/2
Reading, ∆O (inch) x-axis Reading, ∆H (inch) y-axis
1 7.1 1.70 7.1 2.65
2 6.0 1.57 6.2 2.47
3 5.1 1.44 5.3 2.29
4 4.0 1.28 4.3 2.06
5 3.0 1.11 3.0 1.72
By Liner Regression of y on x
Slope, mh = 1.5402 Intercept, ch = 0.0485
*Correction Coefficient, R =  0.9963
Calibration Result: ACCEPT
* If the Correlation Coefficient, R  is < 0.9900. Checking and Recalibration are require.
Remark:  _______________________________________________________________________________________________________________________
Calibrated By:  ______________________________ Date:  ______________________________
Checked By:  _______________________________ Date:  ______________________________
8/7/2000
ARUP TSP - Total Suspended Particulates High Volume Sampler
In-situ Calibration Report (ASR2)
 
Calibration Date 13-May-06 Next Calibration Date 13-Jul-06
Station ASR2 Equipment no. E.HVS.02
    Ambient Condition    
Temperature, Ta (K) 299.6   Pressure, Pa (mmHg) 759.7
    Orifice Transfer Standard Information    
Equipment no. E.CAL.01      
Slope, mo 1.55569 Intercept, co -0.00495
Last Calibration Date 6-Jul-05   Next Calibration Date 6-Jul-06
  mo x Qstd + co = [∆O x (Pa/760) x (298/Ta)]1/2     
     Qstd = {[∆O x (Pa/760) x (298/Ta)]1/2 - co} / mo    
Calibration Point Orifice Manometer   Orifice Qstd (CMM) HVS Manometer  [∆H x (Pa/760) x (298/Ta)]1/2
Reading, ∆O (inch) x-axis Reading, ∆H (inch) y-axis
1 7.1 1.71 7.4 2.71
2 6.0 1.57 6.6 2.56
3 5.1 1.45 5.4 2.32
4 3.9 1.27 4.5 2.12
5 3.0 1.11 3.3 1.81
By Liner Regression of y on x
Slope, mh = 1.4977 Intercept, ch = 0.1717
*Correction Coefficient, R =  0.9951
Calibration Result: ACCEPT
* If the Correlation Coefficient, R  is < 0.9900. Checking and Recalibration are require.
Remark:  ___________________________________________________________________________________________________________
Calibrated By:  ______________________________ Date:  ______________________________
Checked By:  _______________________________ Date:  ______________________________
8/7/2000
ARUP
TSP - Total Suspended Particulates High Volume Sampler
In-situ Calibration Report (ASR3)
 
Calibration Date 4-Jul-06 Next Calibration Date 4-Sep-06
Station ASR2 Equipment no. E.HVS.02
    Ambient Condition    
Temperature, Ta (K) 302.0   Pressure, Pa (mmHg) 754.4
    Orifice Transfer Standard Information    
Equipment no. E.CAL.01      
Slope, mo 1.55569 Intercept, co -0.00495
Last Calibration Date 6-Jul-05   Next Calibration Date 6-Jul-06
  mo x Qstd + co = [∆O x (Pa/760) x (298/Ta)]1/2     
     Qstd = {[∆O x (Pa/760) x (298/Ta)]1/2 - co} / mo    
Calibration Point Orifice Manometer   Orifice Qstd (CMM) HVS Manometer  [∆H x (Pa/760) x (298/Ta)]1/2
Reading, ∆O (inch) x-axis Reading, ∆H (inch) y-axis
1 7.1 1.70 7.2 2.66
2 5.8 1.54 5.9 2.40
3 5.0 1.43 5.1 2.24
4 4.0 1.28 4.1 2.00
5 3.0 1.11 3.0 1.71
By Liner Regression of y on x
Slope, mh = 1.5800 Intercept, ch = -0.0221
*Correction Coefficient, R =  0.9998
Calibration Result: ACCEPT
* If the Correlation Coefficient, R  is < 0.9900. Checking and Recalibration are require.
Remark:  ___________________________________________________________________________________________________________
Calibrated By:  ______________________________ Date:  ______________________________
Checked By:  _______________________________ Date:  ______________________________
8/7/2000
ARUP
TSP - Total Suspended Particulates High Volume Sampler
In-situ Calibration Report (ASR4)
 
Calibration Date 27-Jun-06 Next Calibration Date 27-Aug-06
Station ASR3 Equipment no. E.HVS.03
    Ambient Condition    
Temperature, Ta (K) 302.7   Pressure, Pa (mmHg) 755.3
    Orifice Transfer Standard Information    
Equipment no. E.CAL.01      
Slope, mo 1.55569 Intercept, co -0.00495
Last Calibration Date 6-Jul-05   Next Calibration Date 6-Jul-06
  mo x Qstd + co = [∆O x (Pa/760) x (298/Ta)]1/2     
     Qstd = {[∆O x (Pa/760) x (298/Ta)]1/2 - co} / mo    
Calibration Point Orifice Manometer   Orifice Qstd (CMM) HVS Manometer  [∆H x (Pa/760) x (298/Ta)]1/2
Reading, ∆O (inch) x-axis Reading, ∆H (inch) y-axis
1 6.1 1.57 6.3 2.48
2 5.1 1.44 5.3 2.28
3 4.0 1.27 4.4 2.07
4 3.1 1.12 3.4 1.82
5 2.5 1.01 3.0 1.71
By Liner Regression of y on x
Slope, mh = 1.3797 Intercept, ch = 0.3033
*Correction Coefficient, R =  0.9981
Calibration Result: ACCEPT
* If the Correlation Coefficient, R  is < 0.9900. Checking and Recalibration are require.
Remark:  ___________________________________________________________________________________________________________
Calibrated By:  ______________________________ Date:  ______________________________
Checked By:  _______________________________ Date:  ______________________________