ARUP
TSP - Total Suspended Particulates High Volume Sampler
In-situ Calibration Report
 
Calibration Date 1-Aug-06 Next Calibration Date 1-Oct-06
Station ASR1 Equipment no. E.HVS.01
    Ambient Condition    
Temperature, Ta (K) 302.6   Pressure, Pa (mmHg) 753.6
    Orifice Transfer Standard Information    
Equipment no. E.CAL.01      
Slope, mo 1.56734 Intercept, co -0.02022
Last Calibration Date 25-Jul-06   Next Calibration Date 25-Jul-07
  mo x Qstd + co = [∆O x (Pa/760) x (298/Ta)]1/2     
     Qstd = {[∆O x (Pa/760) x (298/Ta)]1/2 - co} / mo    
Calibration Point Orifice Manometer   Orifice Qstd (CMM) HVS Manometer  [∆H x (Pa/760) x (298/Ta)]1/2
Reading, ∆O (inch) x-axis Reading, ∆H (inch) y-axis
1 7.0 1.68 7.0 2.61
2 6.1 1.57 6.1 2.44
3 5.0 1.42 5.1 2.23
4 4.0 1.27 4.1 2.00
5 3.0 1.10 3.1 1.74
By Liner Regression of y on x
Slope, mh = 1.5117 Intercept, ch = 0.0732
*Correction Coefficient, R =  0.9999
Calibration Result: ACCEPT
* If the Correlation Coefficient, R  is < 0.9900. Checking and Recalibration are require.
Remark:  _______________________________________________________________________________________________________________________
Calibrated By:  ______________________________ Date:  ______________________________
Checked By:  _______________________________ Date:  ______________________________
7/8/2000
ARUP TSP - Total Suspended Particulates High Volume Sampler
In-situ Calibration Report
 
Calibration Date 1-Aug-06 Next Calibration Date 1-Oct-06
Station ASR2 Equipment no. E.HVS.02
    Ambient Condition    
Temperature, Ta (K) 302.5   Pressure, Pa (mmHg) 753.6
    Orifice Transfer Standard Information    
Equipment no. E.CAL.01      
Slope, mo 1.56734 Intercept, co -0.02022
Last Calibration Date 25-Jul-06   Next Calibration Date 25-Jul-07
  mo x Qstd + co = [∆O x (Pa/760) x (298/Ta)]1/2     
     Qstd = {[∆O x (Pa/760) x (298/Ta)]1/2 - co} / mo    
Calibration Point Orifice Manometer   Orifice Qstd (CMM) HVS Manometer  [∆H x (Pa/760) x (298/Ta)]1/2
Reading, ∆O (inch) x-axis Reading, ∆H (inch) y-axis
1 7.0 1.68 7.1 2.63
2 6.0 1.56 6.0 2.42
3 5.1 1.44 5.2 2.25
4 4.2 1.31 4.3 2.05
5 3.0 1.11 3.0 1.71
By Liner Regression of y on x
Slope, mh = 1.5807 Intercept, ch = -0.0262
*Correction Coefficient, R =  0.9995
Calibration Result: ACCEPT
* If the Correlation Coefficient, R  is < 0.9900. Checking and Recalibration are require.
Remark:  ___________________________________________________________________________________________________________
Calibrated By:  ______________________________ Date:  ______________________________
Checked By:  _______________________________ Date:  ______________________________
7/8/2000
ARUP
TSP - Total Suspended Particulates High Volume Sampler
In-situ Calibration Report
 
Calibration Date 1-Aug-06 Next Calibration Date 1-Oct-06
Station ASR3 Equipment no. E.HVS.02
    Ambient Condition    
Temperature, Ta (K) 303.8   Pressure, Pa (mmHg) 753.6
    Orifice Transfer Standard Information    
Equipment no. E.CAL.01      
Slope, mo 1.56734 Intercept, co -0.02022
Last Calibration Date 25-Jul-06   Next Calibration Date 25-Jul-07
  mo x Qstd + co = [∆O x (Pa/760) x (298/Ta)]1/2     
     Qstd = {[∆O x (Pa/760) x (298/Ta)]1/2 - co} / mo    
Calibration Point Orifice Manometer   Orifice Qstd (CMM) HVS Manometer  [∆H x (Pa/760) x (298/Ta)]1/2
Reading, ∆O (inch) x-axis Reading, ∆H (inch) y-axis
1 7.1 1.69 7.3 2.66
2 6.1 1.57 6.2 2.46
3 5.0 1.42 5.2 2.25
4 4.1 1.29 4.3 2.05
5 3.2 1.14 3.4 1.82
By Liner Regression of y on x
Slope, mh = 1.5206 Intercept, ch = 0.0867
*Correction Coefficient, R =  0.9997
Calibration Result: ACCEPT
* If the Correlation Coefficient, R  is < 0.9900. Checking and Recalibration are require.
Remark:  ___________________________________________________________________________________________________________
Calibrated By:  ______________________________ Date:  ______________________________
Checked By:  _______________________________ Date:  ______________________________
7/8/2000